For the timely treatment of patients with infections in bloodstream and cerebrospinal fluid, a rapid antimicrobial susceptibility test (AST) is urgently needed. Here, we describe a direct and rapid antimicrobial susceptibility testing (dRAST) system, which can determine the antimicrobial susceptibility of bacteria from a positive blood culture bottle (PBCB) in six hours. The positive blood culture sample is directly mixed with agarose and inoculated into a micropatterned plastic microchip with lyophilized antibiotic agents. Using microscopic detection of bacterial colony formation in agarose, the total time to result from a PBCB for dRAST was only six hours for a wide range of bacterial concentrations in PBCBs. The results from the dRAST system were consistent with the results from a standard AST, broth microdilution test. In tests of clinical isolates (n = 206) composed of 16 Gram-negative species and seven Gram-positive species, the dRAST system was accurate compared to the standard broth microdilution test, with rates of 91.11% (2613/2868) categorical agreement, 6.69% (192/2868) minor error, 2.72% (50/1837) major error and 1.45% (13/896) very major error. Thus, the dRAST system can be used to rapidly identify appropriate antimicrobial agents for the treatment of blood stream infection (BSI) and antibiotic-resistant strain infections.
CITATION STYLE
Choi, J., Jeong, H. Y., Lee, G. Y., Han, S., Han, S., Jin, B., … Kwon, S. (2017). Direct, rapid antimicrobial susceptibility test from positive blood cultures based on microscopic imaging analysis. Scientific Reports, 7(1). https://doi.org/10.1038/s41598-017-01278-2
Mendeley helps you to discover research relevant for your work.