Endovascular retrieval of a dislocated pushable coil in the common hepatic artery using a cerebral stent retriever

0Citations
Citations of this article
4Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: This is case of removing a dislocated pushable coil from the common hepatic artery (CHA) as a possible complication of using pushable coils in the embolization of an upper gastrointestinal bleeding (UGIB) from the gastroduodenal artery (GDA) by using a pRESET stent retriever (Phenox, Bochum, Germany) which is utilized mainly for treatment of endovascular stroke. Case presentation: An 88-year-old female patient was referred to our hospital to get an emergency embolization of the GDA causing an UGIB with a relevant drop of the hemoglobin level. During the routine embolization of the GDA using pushable coils, a complete dislocation of the last coil into the CHA took place leading to a relevant slowing down of the arterial blood flow to the liver. A decision was thereby made to remove the dislocated coil to avoid further possible complications which was successfully achieved. Conclusions: Various stent retrievers have been proven to be effective in removing dislocated coils during intracerebral coiling of different pathologies. This case report is to our knowledge the first case report proving the high efficacy and safety of using yet another stent retriever, namely a pRESET stent retriever in removing a fully dislocated coil in the abdominal vessels, namely in this case the CHA.

Cite

CITATION STYLE

APA

Shamseldin, M., Stier, A., Hosten, N., & Puls, R. (2021). Endovascular retrieval of a dislocated pushable coil in the common hepatic artery using a cerebral stent retriever. CVIR Endovascular, 4(1). https://doi.org/10.1186/s42155-021-00224-8

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free