FreeStyle navigator continuous glucose monitoring system with TRUstart algorithm, a 1-hour warm-up time

51Citations
Citations of this article
61Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: The first-generation FreeStyle Navigator® Continuous Glucose Monitoring System (FreeStyle Navigator CGM) requires a 10 h warm-up period to avoid inaccurate glucose readings caused by sensor insertion trauma and wound-healing processes. The performance of a second-generation FreeStyle Navigator CGM that begins reporting glucose 1 h after sensor insertion is described. Methods: Second-generation FreeStyle Navigator CGM performance was evaluated in an in-clinic study using the YSI Model 2300 STATPlus Glucose Analyzer as reference with 47 subjects with type 1 diabetes. The reference readings were taken at 15 min intervals, and the study was designed to emphasize the first 10 h of use. Results: The second-generation FreeStyle Navigator CGM exhibited continuous glucose error grid analysis ratings of 93.7% "clinically accurate," 3.6% "benign errors," and 2.8% "clinical errors" and a mean and median absolute relative difference of 14.5% and 10.7%, respectively. The second-generation algorithm detected signal instability in the first 10 h of use and suspended the reporting of 14.1% of first day continuous glucose readings. The clinical accuracy of the second-generation FreeStyle Navigator CGM was similar for the first 10 h versus subsequent hours, with 92.6% and 94.2% "clinically accurate" readings, respectively. Conclusion: The warm-up period for the second-generation FreeStyle Navigator CGM was reduced from 10 to 1 h, with minimal interruption of glucose reporting and without sacrificing clinical performance. © Diabetes Technology Society.

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

McGarraugh, G., Brazg, R., & Weinstein, R. (2011). FreeStyle navigator continuous glucose monitoring system with TRUstart algorithm, a 1-hour warm-up time. Journal of Diabetes Science and Technology, 5(1), 99–106. https://doi.org/10.1177/193229681100500114

Readers over time

‘11‘12‘13‘14‘15‘16‘17‘18‘19‘20‘21‘22‘23‘240481216

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 15

50%

Researcher 12

40%

Professor / Associate Prof. 3

10%

Readers' Discipline

Tooltip

Medicine and Dentistry 14

47%

Engineering 8

27%

Chemistry 4

13%

Nursing and Health Professions 4

13%

Article Metrics

Tooltip
Social Media
Shares, Likes & Comments: 1

Save time finding and organizing research with Mendeley

Sign up for free
0