A validation study of an UV spectrophotometric method was conducted for quantitative analysis of benznidazol (BZN) in a lipid. The analytical determinations were performed at 315 nm at 25 °C. The emulsion components did not interfere on drug analyses, demonstrating the specificity of the methodology. A good linearity (r = 0.99995) and precision (RSD < 5.0%) for intra and inter-day studies, including the reproducibility test were observed. The accuracy ranged of 102.1 ± 2.8 e 103.8 ± 1.7%. The statistical analysis demonstrates a linear, precise, accurate and robust method for BZN quantification from the lipid emulsion system.
CITATION STYLE
Streck, L., Dos Santos, K. S. C. R., De Freitas Fernandes-Pedrosa, M., Da Silva-Júnior, A. A., & De Oliveira, A. G. (2011). Validação de método analítico por espectrofotometria uv para sistema emulsionado lipídico contendo benznidazol. Quimica Nova, 34(8), 1459–1463. https://doi.org/10.1590/S0100-40422011000800028