Evaluation of radiofrequency electronic system in intraoperative monitoring of surgical textiles

1Citations
Citations of this article
15Readers
Mendeley users who have this article in their library.

Abstract

Objective To test performance of SurgiSafe®, a radiofrequency electronic device to detect surgical textiles during operations as compared to manual counting. Methods Surgical sponges with radiofrequency TAGs were placed in the abdominal cavity of a pig submitted to laparotomy, in randomly distributed sites. The TAGs were counted manually and also using SurgiSafe®. Positive and negative predictive values, sensitivity, specificity and time required for counting were analyzed for both methods. Results Through the analysis of 35 surgical cycles, SurgiSafe® immediately identified all sponges, with specificity, sensitivity, positive and negative predictive values of 100%. Although not statistically significant, the manual count had sensitivity of 99.72% and specificity of 99.90%. Conclusion SurgiSafe® proved to be an effective device to identify surgical sponges in vivo, in real time; and its use as an adjuvant to manual counting is very helpful to increase patient's safety.

Cite

CITATION STYLE

APA

Antonio, A. M., & Vieira, C. A. P. (2018). Evaluation of radiofrequency electronic system in intraoperative monitoring of surgical textiles. Einstein (Sao Paulo, Brazil), 16(1), eAO3997. https://doi.org/10.1590/s1679-45082018ao3997