The effects of file size, sodium hypochlorite and blood on the accuracy of Root ZX apex locator in enlarged root canals: An in vitro study

64Citations
Citations of this article
62Readers
Mendeley users who have this article in their library.

Abstract

Background: The initial electronic apex locator (EAL) length measurement is generally established with a small-sized file. It is not known whether file size would be interfering with the reading accuracy of the EAL. This study aimed to evaluate the effect of file size on the accuracy of Root ZX apex locator using an agar model when sodium hypochlorite solution or blood was present during electronic measurements in enlarged root canals. Methods: A total of 36 extracted lower premolars were used. In stage 1, the canals were instrumented using size 10-40 K-files with a size 40 K-file as the master apical file (MAF). The teeth were then divided randomly into two groups of 18 teeth each. In group A, the teeth were mounted in one per cent agar and irrigated with six per cent sodium hypochlorite solution (NaOCl), while in group B the teeth were mounted in agar and irrigated with human blood. In stage 2, the canals were enlarged using a size 60 K-file as the MAF. In stages 1 and 2, the apical portions of the canals were instrumented using the step-back sequence (up to a size 80 K-file). In stage 3, the canals were enlarged using a size 80 K-file as the MAF. In each stage, the length was measured with a Root ZX until the meter value reached 'APEX' using small and large size files. Results: Three-way ANOVA and Bonferroni test showed that file size, stage of preparation and type of irrigant all had a significant influence on the measurement error (P<0.0001), with all the interactions between these three factors being significant (P<0.0001 ). Conclusions: As the diameter of the root canal increased, the measured length with the smaller size files became shorter. A file of a size close to the prepared canal diameter should be used for root length measurement in the presence of blood. In the presence of NaOCl, the Root ZX was highly accurate even when the file was much smaller than the diameter of the canal. The agar model was effective and suitable for testing EALs in vitro.

References Powered by Scopus

Microscopic investigation of root apexes.

396Citations
N/AReaders
Get full text

New Method for Measuring the Length of the Root Canal

208Citations
N/AReaders
Get full text

FACTORS AFFECTING SUCCESSFUL REPAIR AFTER ROOT CANAL THERAPY.

179Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Critical diameter of apical foramen and of file size using the root ZX apex locator: An in vitro study

49Citations
N/AReaders
Get full text

Effectiveness of four electronic apex locators to determine distance from the apical foramen

49Citations
N/AReaders
Get full text

Ex vivo evaluation of the ability of four different electronic apex locators to determine the working length in teeth with various foramen diameters

47Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Ebrahim, A. K., Yoshioka, T., Kobayashi, C., & Suda, H. (2006). The effects of file size, sodium hypochlorite and blood on the accuracy of Root ZX apex locator in enlarged root canals: An in vitro study. Australian Dental Journal, 51(2), 153–157. https://doi.org/10.1111/j.1834-7819.2006.tb00419.x

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 29

83%

Professor / Associate Prof. 4

11%

Lecturer / Post doc 1

3%

Researcher 1

3%

Readers' Discipline

Tooltip

Medicine and Dentistry 36

92%

Decision Sciences 1

3%

Chemistry 1

3%

Engineering 1

3%

Save time finding and organizing research with Mendeley