A liquid chromatographic (LC) method was compared with the AOAC mouse bioassay method (18.086-18.092) for determination of paralytic shellfish toxins in shellfish tissues. Shellfish samples were collected from Massachusetts coastal waters as part of a state surveillance program, and extracts of shellfish meat were analyzed for toxins by using both analytical methods. Overall correlation of the LC and bioassay methods is good (r = 0.943), but for samples with toxicities less than 100 micrograms saxitoxin/100 g shellfish meat, the correlation is significantly less (r = 0.531). Limits of detection are 10 micrograms saxitoxin/100 g shellfish meat and 40 micrograms saxitoxin/100 g shellfish meat for the LC and bioassay methods, respectively. Analytical capacity of the LC method is limited to 12 samples/person-day compared with 30 samples/person-day for the bioassay. Sampling capacity of the LC method could be increased by using a fluorescence detector with a wider response range, which would eliminate the need for dilution of concentrated samples.
CITATION STYLE
Salter, J. E., Timperi, R. J., Hennigan, L. J., Sefton, L., & Reece, H. (1989). Comparison evaluation of liquid chromatographic and bioassay methods of analysis for determination of paralytic shellfish poisons in shellfish tissues. Journal - Association of Official Analytical Chemists, 72(4), 670–673. https://doi.org/10.1093/jaoac/72.4.670