Automation in urinalysis: Evaluation of three urine test strip analysers

6Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A clinical laboratory evaluation was conducted on the Clinitek Auto 2000, the Super aution Analyzer and the Urotron RL9 for the determination of glucose, protein, pH, blood, ketone-bodies and bilirubin. Precision of the systems was tested using three commercial control urine materials, and reported as the percentage of times the instrument repeats a certain value. Good repeatability was obtained with all the instruments. Accuracy of the systems was evaluated by comparison with quantitative procedures, and to check agreement between methods yielding semi-quantitative and quantitative results, ranges of acceptability were defined, based on the criteria reported in a previous paper. It wasx then found that 87.5 to 98.9% of results from the Urotron RL9 and the Clinitek Auto 2000 were acceptable. With the Super Aution Analyzer the level of agreement was apparently lower because of the higher number of concentration steps used by this instrument.

Cite

CITATION STYLE

APA

Bonini, P., Sanguini, L. C., Grossi, L., Ceriotti, F., & Murone, M. (1988). Automation in urinalysis: Evaluation of three urine test strip analysers. Journal of Automatic Chemistry, 10(3), 121–129. https://doi.org/10.1155/S1463924688000239