Safety of localizing epilepsy monitoring intracranial electroencephalograph electrodes using MRI: Radiofrequency-induced heating

71Citations
Citations of this article
112Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: To investigate heating during postimplantation localization of intracranial electroencephalograph (EEG) electrodes by MRI. Materials and Methods: A phantom patient with a realistic arrangement of electrodes was used to simulate tissue heating during MRI. Measurements were performed using 1.5 Tesla (T) and 3T MRI scanners, using head- and body-transmit RF-coils. Two electrode-lead configurations were assessed: a "standard" condition with external electrode-leads physically separated and a "fault" condition with all lead terminations electrically shorted. Results: Using a head-transmit-receive coil and a 2.4 W/kg head-average specific absorption rate (SAR) sequence, at 1.5T the maximum temperature change remained within safe limits (<1°C). Under "standard" conditions, we observed greater heating (≤2.0°C) at 3T on one system and similar heating (<1°C) on a second, compared with the 1.5T system. In all cases these temperature maxima occurred at the grid electrode. In the "fault" condition, larger temperature increases were observed at both field strengths, particularly for the depth electrodes. Conversely, with a body-transmit coil at 3T significant heating (+6.4°C) was observed (same sequence, 1.2/0.5 W/kg head/body-average) at the grid electrode under "standard" conditions, substantially exceeding safe limits. These temperature increases neglect perfusion, a major source of heat dissipation in vivo. Conclusion: MRI for intracranial electrode localization can be performed safely at both 1.5T and 3T provided a head-transmit coil is used, electrode leads are separated, and scanner-reported SARs are limited as determined in advance for specific scanner models, RF coils and implant arrangements. Neglecting these restrictions may result in tissue injury. © 2008 Wiley-Liss, Inc.

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Carmichael, D. W., Thornton, J. S., Rodionov, R., Thornton, R., McEvoy, A., Allen, P. J., & Lemieux, L. (2008). Safety of localizing epilepsy monitoring intracranial electroencephalograph electrodes using MRI: Radiofrequency-induced heating. Journal of Magnetic Resonance Imaging, 28(5), 1233–1244. https://doi.org/10.1002/jmri.21583

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 40

51%

Researcher 22

28%

Professor / Associate Prof. 15

19%

Lecturer / Post doc 2

3%

Readers' Discipline

Tooltip

Medicine and Dentistry 33

43%

Engineering 18

24%

Neuroscience 16

21%

Agricultural and Biological Sciences 9

12%

Article Metrics

Tooltip
Mentions
Blog Mentions: 1

Save time finding and organizing research with Mendeley

Sign up for free