We conducted a comparative evaluation of the Chlamydiazyme (Abbott Laboratories), PACE 2 (Gen-Probe), and AMP-CT (Gen-Probe) assays for the detection of Chlamydia trachomatis in endocervical samples. Specimens from 787 females were included in the study. The sensitivities of the PACE 2 and Chlamydiazyme assays in comparison to the results of the AMP-CT assay were 79.3 and 63.4%, respectively. The specificities of the Chlamydiazyme and PACE 2 assays were 100%. All of the positive specimens detected in this study were positive by the AMP-CT assay. On the basis of the final results of the comparison, the prevalence of C. trachomatis in the population was 10.4%. Retesting of specimens whose results were in the intermediate zone by the PACE 2 assay by a probe competition assay identified some additional true- positive specimens. Amplification assay testing of such specimens did not significantly increase the yield. The majority of specimens which tested positive by the AMP-CT assay only were not in the intermediate zone by the PACE 2 assay. We were unable to identify demographic or clinical factors which could predict those individuals who tested positive by amplified tests but not by nonamplified tests. The Gen-Probe PACE 2 assay proved to be superior to the Chlamydiazyme assay for the screening and diagnosis of C. trachomatis infections in female endocervical specimens.
CITATION STYLE
Wylie, J. L., Moses, S., Babcock, R., Jolly, A., Giercke, S., & Hammond, G. (1998). Comparative evaluation of Chlamydiazyme, PACE 2, and AMP-CT assays for detection of Chlamydia trachomatis in endocervical specimens. Journal of Clinical Microbiology, 36(12), 3488–3491. https://doi.org/10.1128/jcm.36.12.3488-3491.1998
Mendeley helps you to discover research relevant for your work.