Feasibility of Xpert Ebola assay in médecins sans Frontières Ebola program, Guinea

20Citations
Citations of this article
65Readers
Mendeley users who have this article in their library.

Abstract

Rapid diagnostic methods are essential in control of Ebola outbreaks and lead to timely isolation of cases and improved epidemiologic surveillance. Diagnosis during Ebola outbreaks in West Africa has relied on PCR performed in laboratories outside this region. Because time between sampling and PCR results can be considerable, we assessed the feasibility and added value of using the Xpert Ebola Assay in an Ebola control program in Guinea. A total of 218 samples were collected during diagnosis, treatment, and convalescence of patients. Median time for obtaining results was reduced from 334 min to 165 min. Twenty-six samples were positive for Ebola virus. Xpert cycle thresholds were consistently lower, and 8 (31%) samples were negative by routine PCR. Several logistic and safety issues were identified. We suggest that implementation of the Xpert Ebola Assay under programmatic conditions is feasible and represents a major advance in diagnosis of Ebola virus disease without apparent loss of assay sensitivity.

Cite

CITATION STYLE

APA

Van Den Bergh, R., Chaillet, P., Sow, M. S., Amand, M., Van Vyve, C., Jonckheere, S., … Antierens, A. (2016). Feasibility of Xpert Ebola assay in médecins sans Frontières Ebola program, Guinea. Emerging Infectious Diseases, 22(2), 210–216. https://doi.org/10.3201/eid2202.151238