The current technology for monitoring fetal wellbeing during child birth is cardiotocography. However, CTG has high false positive rates that lead to unnecessary emergency Cesarean deliveries and false negatives that result in birth injuries. To curtail these issues, fetal pulse oximetery has been a topic of interest for many decades. Fetal pulse oximetry would yield the oxygen saturation of the fetus in utero and provide a more robust marker for clinicians to make decisions about performing emergency Cesarean deliveries. Here, we present a review of biomedical optical developments related to transabdominal fetal pulse oximetery in the biophotonics field and the challenges that must be overcome to make transabdominal pulse oximetry a clinical reality.
CITATION STYLE
Gunther, J. E., Jayet, B., Sekar, S. K. V., Kainerstorfer, J. M., & Andersson-Engels, S. (2022, June 1). Review of optical methods for fetal monitoring in utero. Journal of Biophotonics. John Wiley and Sons Inc. https://doi.org/10.1002/jbio.202100343