It was performed an analysis of the uncertainties involved in a cross-calibration procedure of an ionization chamber used in Sírio-Libanês Hospital, according to the methodology proposed in IAEA-TECDOC-1585, based on ISO GUM. The overall uncertainty obtained was 2,55% (coverage factor k = 2), which meant an increase of 0,04% compared to the reference ionization chamber uncertainty reported by the dosimetry laboratory of LCI-GMR-IPEN/CNEN-SP.
CITATION STYLE
Cardoso, P. H. B., Tietz, G. F., Neves-Junior, W. F. P., Haddad, C. M. K., & Terini, R. A. (2015). Determination of the uncertainty in the cross-calibration of an ionization chamber used in radiation therapy. In IFMBE Proceedings (Vol. 51, pp. 756–758). Springer Verlag. https://doi.org/10.1007/978-3-319-19387-8_186