NIST calibration of a neutron spectrometer ROSPEC

7Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.

Abstract

A neutron spectrometer was acquired for use in the measurement of National Institute of Standards and Technology neutron fields. The spectrometer included options for the measurement of low and high energy neutrons, for a total measurement range from 0.01 eV up to 17 MeV. The spectrometer was evaluated in calibration fields and was used to determine the neutron spectrum of an Americium-Beryllium neutron source. The calibration fields used included bare and moderated 252Cf, monoenergetic neutron fields of 2.5 MeV and 14 MeV, and a thermal-neutron beam. Using the calibration values determined in this exercise, the spectrometer gives a good approximation of the neutron spectrum, and excellent values for neutron fluence, for all NIST calibration fields. The spectrometer also measured an Americium-Beryllium neutron field in a NIST exposure facility and determined the field quite well. The spectrometer measured scattering effects in neutron spectra which previously could be determined only by calculation or integral measurements.

Cite

CITATION STYLE

APA

Heimbach, C. (2006). NIST calibration of a neutron spectrometer ROSPEC. Journal of Research of the National Institute of Standards and Technology, 111(6), 419–428. https://doi.org/10.6028/jres.111.032