A pH sensitivity control method using hydrogen annealing for reference field-effect transistors

0Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

Many studies have been carried out on the use of ion-sensitive field-effect transistor (ISFET) and reference FET (REFET) devices for pH measurements. In previous works, different materials were used for the sensing membranes of ISFET and REFET devices so that the pH sensitivity of the REFET was much less than that of the ISFET. In these cases, it proved difficult to cancel out the variation in the AC voltage component in the solution by synchronizing the output signals of the REFET and ISFET. Therefore, we put forward a new proposal, in which we use hydrogen annealing to decrease the pH sensitivity of a Si 3N4 film. Devices comprising Si3N4 on SiO2 on n-type Si substrates were fabricated. The pH sensitivity of a device annealed in hydrogen for 20 min at 400 °C was reduced from 54.6 to 32.6 mV/pH. A higher temperature annealing was largely ineffective in producing greater decreases in sensitivity, and the annealing affected only the surface of the Si3N4 film, not its interior. We succeeded in developing the fabrication technology for a device with low pH sensitivity to be used as a REFET.

Cite

CITATION STYLE

APA

Futagawa, M., Takahashi, M., Kamado, K., Ishida, M., & Sawada, K. (2013). A pH sensitivity control method using hydrogen annealing for reference field-effect transistors. Sensors and Materials, 25(9), 673–688. https://doi.org/10.18494/sam.2013.917

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free