Well-defined quality control (QC) processes are used to determine whether a certain procedure or action conforms to a widely accepted standard and/or set of guidelines, and are important components of any laboratory quality assurance program (Popa-Burke et al., J Biomol Screen 14: 1017–1030, 2009). In this chapter, we describe QC procedures useful for monitoring the accuracy and precision of laboratory instrumentation, most notably automated liquid dispensers. Two techniques, gravimetric QC and photometric QC, are highlighted in this chapter. When used together, these simple techniques provide a robust process for evaluating liquid handler accuracy and precision, and critically underpin high-quality research programs.
CITATION STYLE
Jayakody, C., & Hull-Ryde, E. A. (2016). Instrument quality control. In Methods in Molecular Biology (Vol. 1439, pp. 99–113). Humana Press Inc. https://doi.org/10.1007/978-1-4939-3673-1_6