Optimizing solute removal at minimized albumin loss is a major goal of dialyzer engineering. In a prospective, randomized, crossover study on eight patients (age 63 ± 14 years) on maintenance hemodialysis, the new Baxter Xenium 170 high-flux dialyzer (BX), which contains a 1.7-m2 PUREMA H dialysis membrane, was compared with two widely used reference high-flux dialyzers currently available for hemodialysis in North America, the Fresenius Optiflux 180 NR (FO) and the Gambro Polyflux 170 H (GP). Solute removal and biocompatibility were assessed in hemodialysis for 240 min at blood and dialysate flow rates of 300 and 500 mL/min, respectively. Additional ex vivo experiments detecting the interleukin-1beta (IL-1b) generation in recirculated donor blood were performed to demonstrate the pyrogen retention properties of the dialyzers. The instantaneous plasma clearances were similar for the three dialyzers except for cystatin c (cysc), for which a lower clearance was measured with FO as compared with BX and GP after 30 and 180 min of hemodialysis. The reduction ratios (RRs) corrected for the hemoconcentration of beta 2-microglobulin and cysc were lower in FO (44 ± 9 and 35 ± 9%, respectively) versus BX (62 ± 6 and 59 ± 7%, respectively) and GP (61 ± 7 and 56 ± 8%, respectively). The RRs of the cytokine tumor necrosis factor alpha and interleukin-6 were not different between the dialyzers. The albumin loss was <300 mg for all filters. No differences between the dialyzers were found in the biocompatibility parameters showing very low leukocyte and complement activation. The ex vivo recirculation experiments revealed a significantly higher IL-1b generation for GP (710 ± 585 pg/mL) versus BX (317 ± 211 pg/mL) and FO (151 ± 38 pg/mL). BX is characterized by a steep solute sieving profile with high low-molecular weight protein removal at virtually no albumin loss and an excellent biocompatibility. This improved performance may be regarded as a contribution to optimal dialysis therapy. © 2008, Copyright the Authors.

Cite

CITATION STYLE

APA

Krieter, D. H., Lemke, H. D., & Wanner, C. (2008). A new synthetic dialyzer with advanced permselectivity for enhanced low-molecular weight protein removal. Artificial Organs, 32(7), 547–554. https://doi.org/10.1111/j.1525-1594.2008.00583.x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free