Report of AAPM Task Group 219 on independent calculation-based dose/MU verification for IMRT

49Citations
Citations of this article
115Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Independent verification of the dose per monitor unit (MU) to deliver the prescribed dose to a patient has been a mainstay of radiation oncology quality assurance (QA). We discuss the role of secondary dose/MU calculation programs as part of a comprehensive QA program. This report provides guidelines on calculation-based dose/MU verification for intensity modulated radiation therapy (IMRT) or volumetric modulated arc therapy (VMAT) provided by various modalities. We provide a review of various algorithms for “independent/second check” of monitor unit calculations for IMRT/VMAT. The report makes recommendations on the clinical implementation of secondary dose/MU calculation programs; on commissioning and acceptance of various commercially available secondary dose/MU calculation programs; on benchmark QA and periodic QA; and on clinically reasonable action levels for agreement of secondary dose/MU calculation programs.

Cite

CITATION STYLE

APA

Zhu, T. C., Stathakis, S., Clark, J. R., Feng, W., Georg, D., Holmes, S. M., … Xiao, Y. (2021). Report of AAPM Task Group 219 on independent calculation-based dose/MU verification for IMRT. Medical Physics, 48(10), e808–e829. https://doi.org/10.1002/mp.15069