The relationship between ultrasonic parameters (attenuation coefficients and velocity) and bone porosity in bovine cancellous bone is explored to understand the possibility of fracture risk diagnosis associated with osteoporosis by applying ultrasound. In vitro measurements of ultrasonic parameters on twenty-one bovine cancellous bone samples from tibia were conducted, using ultrasonic spectroscopy in the through-transmission mode. Transducers of three different center frequencies were used to cover a wide diagnostic frequency range between 1.0–7.8 MHz. The nonlinear relationship of porosity and normalized attenuation coefficient (nATTN) and normalized broadband attenuation coefficient (nBUA) were well described by a third-order polynomial fit, whereas porosity and the phase velocity (UV) were found to be negatively correlated with the linear correlation coefficients of −0.93, −0.89 and −0.83 at 2.25, 5.00 and 7.50 MHz, respectively. The results imply that the ultrasound parameters attain maximum values for the bone sample with the lowest porosity, and then decrease for samples with greater porosity for the range of porosities in our samples for all frequencies. Spatial variation in the ultrasound parameters was found to be caused by non-uniform pore size distribution, which was examined at five different locations within the same bone specimen. However, it did not affect the relationship of ultrasound parameters and porosity at these frequencies.
CITATION STYLE
Karki, A., & Wu, J. (2022). Measurement of Ultrasound Parameters of Bovine Cancellous Bone as a Function of Frequency for a Range of Porosities via Through-Transmission Ultrasonic Spectroscopy. Acoustics, 4(2), 406–418. https://doi.org/10.3390/ACOUSTICS4020025
Mendeley helps you to discover research relevant for your work.