A prototype high-resolution Small-Animal PET scanner dedicated to mouse brain imaging

102Citations
Citations of this article
61Readers
Mendeley users who have this article in their library.

Abstract

We developed a prototype small-animal PET scanner based on depth-encoding detectors using dual-ended readout of small scintillator elements to produce high and uniform spatial resolution suitable for imaging the mouse brain. Methods: The scanner consists of 16 tapered dual-ended-readout detectors arranged in a 61-mm-diameter ring. The axial field of view (FOV) is 7 mm, and the transaxial FOV is 30 mm. The scintillator arrays consist of 14 × 14 lutetium oxyorthosilicate elements, with a crystal size of 0.43 × 0.43 mm at the front end and 0.80 × 0.43 mm at the back end, and the crystal elements are 13 mm long. The arrays are read out by 8 × 8 mm and 13 × 8 mm position-sensitive avalanche photodiodes (PSAPDs) placed at opposite ends of the array. Standard nuclear-instrumentation-module electronics and a custom-designed multiplexer are used for signal processing. Results: The detector performance was measured, and all but the crystals at the very edge could be clearly resolved. The average intrinsic spatial resolution in the axial direction was 0.61 mm. A depth-of-interaction resolution of 1.7 mm was achieved. The sensitivity of the scanner at the center of the FOV was 1.02% for a lower energy threshold of 150 keV and 0.68% for a lower energy threshold of 250 keV. The spatial resolution within a FOV that can accommodate the entire mouse brain was approximately 0.6 mm using a 3-dimensional maximum-likelihood expectation maximization reconstruction. Images of a hot-rod microphantom showed that rods with a diameter of as low as 0.5 mm could be resolved. The first in vivo studies were performed using 18F-fluoride and confirmed that a 0.6-mm resolution can be achieved in the mouse head in vivo. Brain imaging studies with 18F-FDG were also performed. Conclusion: We developed a prototype PET scanner that can achieve a spatial resolution approaching the physical limits of a small-bore PET scanner set by positron range and detector interaction. We plan to add more detector rings to extend the axial FOV of the scanner and increase sensitivity.

References Powered by Scopus

GATE: A simulation toolkit for PET and SPECT

1891Citations
N/AReaders
Get full text

Fast calculation of the exact radiological path for a three dimensional CT array

1378Citations
N/AReaders
Get full text

Fundamental limits of spatial resolution in PET

417Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Innovations in Instrumentation for Positron Emission Tomography

94Citations
N/AReaders
Get full text

Machine learning in quantitative PET: A review of attenuation correction and low-count image reconstruction methods

84Citations
N/AReaders
Get full text

Small animal PET: A review of what we have done and where we are going

53Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Yang, Y., Bec, J., Zhou, J., Zhang, M., Judenhofer, M. S., Bai, X., … Cherry, S. R. (2016). A prototype high-resolution Small-Animal PET scanner dedicated to mouse brain imaging. Journal of Nuclear Medicine, 57(7), 1130–1135. https://doi.org/10.2967/jnumed.115.165886

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 30

67%

Researcher 11

24%

Professor / Associate Prof. 4

9%

Readers' Discipline

Tooltip

Engineering 18

46%

Physics and Astronomy 15

38%

Computer Science 4

10%

Neuroscience 2

5%

Save time finding and organizing research with Mendeley

Sign up for free