Variation in arterial input function in a large multicenter computed tomography perfusion study

9Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Objectives: To report the variation in computed tomography perfusion (CTP) arterial input function (AIF) in a multicenter stroke study and to assess the impact this has on CTP results. Methods: CTP datasets from 14 different centers were included from the DUtch acute STroke (DUST) study. The AIF was taken as a direct measure to characterize contrast bolus injection. Statistical analysis was applied to evaluate differences in amplitude, area under the curve (AUC), bolus arrival time (BAT), and time to peak (TTP). To assess the clinical relevance of differences in AIF, CTP acquisitions were simulated with a realistic anthropomorphic digital phantom. Perfusion parameters were extracted by CTP analysis using commercial software (IntelliSpace Portal (ISP), version 10.1) as well as an in-house method based on block-circulant singular value decomposition (bSVD). Results: A total of 1422 CTP datasets were included, ranging from 6 to 322 included patients per center. The measured values of the parameters used to characterize the AIF differed significantly with approximate interquartile ranges of 200–750 HU for the amplitude, 2500–10,000 HU·s for the AUC, 0–17 s for the BAT, and 10–26 s for the TTP. Mean infarct volumes of the phantom were significantly different between centers for both methods of perfusion analysis. Conclusions: Although guidelines for the acquisition protocol are often provided for centers participating in a multicenter study, contrast medium injection protocols still vary. The resulting volumetric differences in infarct core and penumbra may impact clinical decision making in stroke diagnosis. Key Points: • The contrast medium injection protocol may be different between stroke centers participating in a harmonized multicenter study. • The contrast medium injection protocol influences the results of X-ray computed tomography perfusion imaging. • The contrast medium injection protocol can impact stroke diagnosis and patient selection for treatment.

Cite

CITATION STYLE

APA

Peerlings, D., Bennink, E., Dankbaar, J. W., Velthuis, B. K., de Jong, H. W. A. M., Majoie, C. B., … van der Graaf, Y. (2021). Variation in arterial input function in a large multicenter computed tomography perfusion study. European Radiology, 31(11), 8317–8325. https://doi.org/10.1007/s00330-021-08067-6

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free