To verify the fluence distribution and number of monitor units for dynamic treatment techniques each patient treatment plan is applied to a measuring device before first treatment. We verify Rapid Arc treatment plans patient oriented using the IMRT-MatriXX (IBA, Schwarzenbruck, Germany). Therefore we generated a verification protocol. 3 patients with cerebral metastasis, 2 patients with head and neck cancer, and 5 patients with prostate cancer receiving Rapid Arc radiotherapy were included in this study. It is poss-ible to verify the 2D fluence distribution and the number of monitor units of Rapid Arc treatment plans using IMRT-MatriXX within a few minutes. Therefore we recommend the use of the IMRT-MatriXX for Rapid Arc treatment plans in the clinical routine. © 2009 Springer-Verlag.
CITATION STYLE
Wagner, D. M., & Vorwerk, H. (2009). First experience with patient oriented Rapid Arc treatment plan verification. In IFMBE Proceedings (Vol. 25, pp. 338–340). Springer Verlag. https://doi.org/10.1007/978-3-642-03474-9_95