In this paper we demonstrate significant improvements to the stability of the monochromatic X-ray beam intensity on several beamlines at Diamond, using a modulation of the pitch axis of the DCM with a piezoelectric actuator. The modulation is detected on an intensity diagnostic (e.g. an ion chamber) using a software lock-in technique. The detected amplitude and phase are used in a feedback to keep the DCM at the peak of the rocking curve, or any arbitrary position 'off-peak' which might be desired to detune the DCM and reject unwanted harmonics. A major advantage of this software based system is the great flexibility offered, using standard, readily available instrumentation. Measurements of the short and long-term performance of the feedback on several beamlines are presented, and the limitations of such a feedback are discussed. © 2013 IOP Publishing Ltd.
CITATION STYLE
Bloomer, C., Dent, A., Diaz-Moreno, S., Dolbnya, I., Pedersen, U., Rehm, G., … Thomas, C. (2013). Using DCM pitch modulation and feedback to improve long term X-ray beam stability. In Journal of Physics: Conference Series (Vol. 425). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/425/4/042010