The suppression of the probe laser pumping effect in SERF atomic magnetometer

0Citations
Citations of this article
1Readers
Mendeley users who have this article in their library.

Abstract

Introduction: A spin exchange relaxation free atomic magnetometer, as an ultra-highly sensitive magnetic field sensor, is limited by the performance of the probe laser system. The probe laser pumping effect (PLPE) hinders the increase in the performance of probe laser system. Methods: This study investigated the PLPE and proposed a method for suppressing the same. Through changes to the angle of a quarter wave plate and the addition of a triangular modulated magnetic field to the alkali atoms, the suppression point was determined. Results and discussion: Further, related parameters were measured for different degree of polarizations of the probe laser, which confirmed that the influence of PLPE on the magnetic field was the least at the suppressed point.

Cite

CITATION STYLE

APA

Cao, Q., Zhai, Y., Huang, B., Li, R., & Li, J. (2023). The suppression of the probe laser pumping effect in SERF atomic magnetometer. Frontiers in Physics, 11. https://doi.org/10.3389/fphy.2023.1181399

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free