Validation of the integration of technology that measures additional "vascular" indices into an ambulatory blood pressure monitoring system

19Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.

Abstract

Background: The objective of this study was to validate the novel integration of oscillometric (Vasotens®) technology into a BPLab® ambulatory blood pressure (BP) monitoring system to measure central BP, the aortic augmentation index, and pulse wave velocity (PWV) compared with the recommended and widely accepted tonometric method. Methods: The ARTERY Society guidelines for comparison of PWV measurement techniques were used as the basis for recruitment of 99 individuals (mean age 44±19 years, 52 males). The standard for comparison was the conventional "classic" SphygmoCor device. Results: Accordance of the two methods was satisfactory (r=0.98, mean difference of 2.9±3.5 mmHg for central systolic BP; r=0.98, mean difference of -1.1±2.3 mmHg for central diastolic BP; r=0.83, mean difference of -2.6%±13% for aortic augmentation index; r=0.85, mean difference of 0.69±1.4 for PWV). Conclusion: The performance of Vasotens algorithms using an oscillometric ambulatory BP monitoring system is feasible for accurate diagnosis, risk assessment, and evaluation of the effects of antihypertensive drugs. © 2014 Kotovskaya et al.

Cite

CITATION STYLE

APA

Kotovskaya, Y. V., Kobalava, Z. D., & Orlov, A. V. (2014). Validation of the integration of technology that measures additional “vascular” indices into an ambulatory blood pressure monitoring system. Medical Devices: Evidence and Research, 7(1), 91–97. https://doi.org/10.2147/MDER.S61839

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free