An instrumental set up including on-line solid-phase extraction, nano-liquid chromatography, and nanospray mass spectrometry is constructed to improve the sensitivity for quantitation of perfluorooctanoic acid (PFOA) and perfluorooctane sulfonate (PFOS) in surface water. Sample volumes of 1000 μL are loaded onto a microbore 1.0-mm i.d. x 5 mm, 5 μm Kromasil C18 enrichment column by a carrier solution consisting of 10mM ammonium acetate in acetonitrile-water (10:90, v/v) at a flow rate of 250 μL/min, providing on-line analyte enrichment and sample clean-up. Backflushed elution onto a 0.1-mm i.d. x 150 mm, 3.5 μm Kromasil C18 analytical column is conducted using an acetonitrile-10mM ammonium acetate solvent gradient from 30% to 70% acetonitrile. Water samples are added with internal standard (perfluoroheptanoic acid) and filtrated prior to injection. The mass limits of detection of PFOA and PFOS are 0.5 and 1 pg, respectively, corresponding to concentration limits of detection of 500 pg/L and 1 ng/L, respectively. The total time spent on sample preparation, chromatography, and detection is approximately 12 min per sample. The method was employed for the determination of PFOS and PFOA in urban river water.
CITATION STYLE
Wilson, S. R., Malerød, H., Holm, A., Molander, P., Lundanes, E., & Greibrokk, T. (2007). On-line SPE-nano-LC-nanospray-MS for rapid and sensitive determination of perfluorooctanoic acid and perfluorooctane sulfonate in river water. Journal of Chromatographic Science, 45(3), 146–152. https://doi.org/10.1093/chromsci/45.3.146
Mendeley helps you to discover research relevant for your work.