We compared the Bonfils™ and SensaScope™ rigid fibreoptic scopes in 200 patients with a simulated difficult airway randomised to one of the two devices. A cervical collar inhibited neck movement and reduced mouth opening to a mean (SD) of 23 (3) mm. The primary outcome parameter was overall success of tracheal intubation; secondary outcomes included first-attempt success, intubation times, difficulty of intubation, fibreoptic view and side-effects. The mean (95% CI) overall success rate was 88 (80–94)% for the Bonfils and 89 (81–94)% for the SensaScope (p = 0.83). First-attempt intubation success rates were 63 (53–72)% for the Bonfils and 72 (62–81)% for the SensaScope (p = 0.17). Median (IQR [range]) intubation time was significantly shorter with the SensaScope (34 (20–84 [5–240]) s vs. 45 (25–134 [12–230]) s), and fibreoptic view was significantly better with the SensaScope (full view of the glottis in 79% with the SensaScope vs. 61% with the Bonfils). This might be explained by its steerable tip and the S-formed shape, contributing to better manoeuvrability. There were no differences in the difficulty of intubation or side-effects.
CITATION STYLE
Kleine-Brueggeney, M., Greif, R., Urwyler, N., Wirthmüller, B., & Theiler, L. (2016). The performance of rigid scopes for tracheal intubation: a randomised, controlled trial in patients with a simulated difficult airway. Anaesthesia, 71(12), 1456–1463. https://doi.org/10.1111/anae.13626