Evaluation of the Ultrastructural and In Vitro Flow Properties of the PRESERFLO MicroShunt

16Citations
Citations of this article
26Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Purpose: To measure the in vitro flow properties of the PRESERFLO implant for comparison with the theoretical resistance to flow. Methods: The PRESERFLO was designed to control the flow of aqueous humor according to the Hagen-Poiseuille (HP) equation. Scanning electron microscopy (SEM) was performed to analyze the ultrastructure, and flow measurements were carried out using a gravity-flow setup. Results: SEM images of the PRESERFLO showed luminal diameters of 67.73 × 65.95 μm and 63.66 × 70.54 μm. The total diameter was 337.2 μm, and the wall was 154 μm wide. The theoretical calculation of the resistance to flow (R) for an aqueous humor (AH) viscosity of 0.7185 centipoises (cP) was 1.3 mm Hg/(μL/min). Hence, assuming a constant AH flow of 2 μL/min, the pressure differential across the device (ΔP) was estimated to be 2.6 mm Hg. The gravity-flow experiment allowed us to measure the experimental resistance to flow, which was RE = 1.301 mm Hg/(μL/min), in agreement with the theoretical resistance to flow R given by the HP equation. Conclusions: The experimental and theoretical flow testing showed that the pressure drop across this device would not be large enough to avoid hypotony unless the resistance to outflow of the sub-Tenon space was sufficient to control the intraocular pressure in the early postoperative period. Translational Relevance: The fluid properties of glaucoma subconjunctival drainage devices determine their specific bleb-forming capacity and ability to avoid hypotony and therefore their safety and efficacy profile.

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Barberá, M. I., Hernández-Verdejo, J. L., Bragard, J., Burguete, J., Fernández, L. M., Rivero, P. T., … Teus, M. A. (2021). Evaluation of the Ultrastructural and In Vitro Flow Properties of the PRESERFLO MicroShunt. Translational Vision Science and Technology, 10(13). https://doi.org/10.1167/tvst.10.13.26

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 6

75%

Professor / Associate Prof. 1

13%

Researcher 1

13%

Readers' Discipline

Tooltip

Medicine and Dentistry 6

60%

Business, Management and Accounting 2

20%

Engineering 1

10%

Neuroscience 1

10%

Save time finding and organizing research with Mendeley

Sign up for free