Hydroxyapatite-coated femoral stems

110Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.

Abstract

We performed a histological and histomorphometric examination in five cadaver specimens of the femoral and acetabular components and the associated tissue which had been recovered between 3.3 and 6.2 years after primary total hip arthroplasty (THA) using a proximal hydroxyapatite (HA)-coated titanium alloy implant. All had functioned well during the patients' life. All the stems were fixed in the femur and showed osseointegration of both the proximal and distal parts. The amount of residual HA was greatest in the distal metaphyseal sections, indicating that the rate of bone remodelling may be the main factor causing loss of HA. The level of activity of the patient was the only clinical factor which correlated with loss of coating. The percentage of bone-implant osseointegration was almost constant, regardless of the amount of HA residue, periprosthetic bone density or the time of implantation. HA debris was seldom observed and if present did not cause any adverse or inflammatory reaction. Partial debonding did occur in one case as a result of a polyethylene-induced inflammatory reaction.

Cite

CITATION STYLE

APA

Tonino, A. J., Thèrin, M., & Doyle, C. (1999). Hydroxyapatite-coated femoral stems. Journal of Bone and Joint Surgery - Series B, 81(1), 148–154. https://doi.org/10.1302/0301-620X.81B1.8948