Remote stimulation of sciatic nerve using cuff electrodes and implanted diodes

5Citations
Citations of this article
30Readers
Mendeley users who have this article in their library.

Abstract

We demonstrate a method of neurostimulation using implanted, free-floating, inter-neural diodes. They are activated by volume-conducted, high frequency, alternating current (AC) fields and address the issue of instability caused by interconnect wires in chronic nerve stimulation. The aim of this study is to optimize the set of AC electrical parameters and the diode features to achieve wireless neurostimulation. Three different packaged Schottky diodes (1.5 mm, 500 μm and 220 μm feature sizes) were tested in vivo (n = 17 rats). A careful assessment of sciatic nerve activation as a function of diode-dipole lengths and relative position of the diode was conducted. Subsequently, free-floating Schottky microdiodes were implanted in the nerve (n = 3 rats) and stimulated wirelessly. Thresholds for muscle twitch responses increased non-linearly with frequency. Currents through implanted diodes within the nerve suffer large attenuations (~100 fold) requiring 1-2 mA drive currents for thresholds at 17 μA. The muscle recruitment response using electromyograms (EMGs) is intrinsically steep for subepineurial implants and becomes steeper as diode is implanted at increasing depths away from external AC stimulating electrodes. The study demonstrates the feasibility of activating remote, untethered, implanted microscale diodes using external AC fields and achieving neurostimulation.

References Powered by Scopus

Wireless power transfer to deep-tissue microimplants

419Citations
N/AReaders
Get full text

Energy harvesting for the implantable biomedical devices: Issues and challenges

252Citations
N/AReaders
Get full text

Stimulus Waveforms for Selective Neural Stimulation

219Citations
N/AReaders
Get full text

Cited by Powered by Scopus

A novel microwire interface for small diameter peripheral nerves in a chronic, awake murine model

13Citations
N/AReaders
Get full text

Rapid and Low Cost Manufacturing of Cuff Electrodes

7Citations
N/AReaders
Get full text

Editorial for the special issue on neural electrodes: Design and applications

3Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Sridharan, A., Chirania, S., Towe, B. C., & Muthuswamy, J. (2018). Remote stimulation of sciatic nerve using cuff electrodes and implanted diodes. Micromachines, 9(11). https://doi.org/10.3390/mi9110595

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 9

64%

Researcher 5

36%

Readers' Discipline

Tooltip

Engineering 8

73%

Pharmacology, Toxicology and Pharmaceut... 1

9%

Chemistry 1

9%

Physics and Astronomy 1

9%

Save time finding and organizing research with Mendeley

Sign up for free