How effective are cardiopulmonary bypass circuits at removing gaseous microemboli?

66Citations
Citations of this article
16Readers
Mendeley users who have this article in their library.

Abstract

An association has been demonstrated between intravascular microemboli and organ injury during cardiopulmonary bypass (CPB). Air may be inadvertently introduced into the venous line during CPB resulting in the formation of gaseous microemboli (GME). We studied the ability of CPB circuits, from five different manufacturers, to remove GME originating from the introduction of air into the venous line. Using an in vitro model of adult CPB, 60 ml of air was introduced into the venous line and the progression of GME through the circuit components was monitored at 5 locations. In all circuits GME were detected in the arterial line following the introduction of air into the venous line. There was a wide variation between manufacturers in the ability of the circuit to remove GME. Air introduced into the venous line during CPB results in the formation of GME that are able to pass through all the circuit components including the arterial filter. The quantity of GME detected in the arterial line is influenced by the design of the circuit components and varies between manufacturers. Air in the venous line should be avoided and if present it must be dealt with promptly.

References Powered by Scopus

The impact of microemboli during cardiopulmonary bypass on neuropsychological functioning

674Citations
N/AReaders
Get full text

Cerebral emboli and cognitive outcome after cardiac surgery

226Citations
N/AReaders
Get full text

Cerebral microemboli during cardiopulmonary bypass: Increased emboli during perfusionist interventions

166Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Microbubbles: Pathophysiology and clinical implications

263Citations
N/AReaders
Get full text

Cardiopulmonary bypass management and neurologic outcomes: An evidence-based appraisal of current practices

250Citations
N/AReaders
Get full text

Experimental study of bubble formation at metal porous spargers: Effect of liquid properties and sparger characteristics on the initial bubble size distribution

146Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Jones, T. J., Deal, D. D., Vernon, J. C., Blackburn, N., & Stump, D. A. (2002). How effective are cardiopulmonary bypass circuits at removing gaseous microemboli? Journal of Extra-Corporeal Technology, 34(1), 34–39. https://doi.org/10.1051/ject/200234134

Readers over time

‘11‘12‘13‘15‘16‘17‘18‘19‘20‘21‘2401234

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 6

50%

Professor / Associate Prof. 3

25%

Researcher 3

25%

Readers' Discipline

Tooltip

Medicine and Dentistry 10

77%

Materials Science 1

8%

Nursing and Health Professions 1

8%

Engineering 1

8%

Save time finding and organizing research with Mendeley

Sign up for free
0