Determination of kQ using MLC-collimated rectangular fields for absolute dosimetry of the CyberKnife

1Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Traditional CyberKnife (CK) calibration uses TG-51, which requires kQ to be defined using the standard reference condition of 100 cm SSD in a 10 cm × 10 cm field. Since the CK is calibrated using a 6 cm fixed-aperture collimating cone at 80 cm SAD, the BJR-25 method is commonly used to relate circular-field PDDs to square-field PDDs for kQ determination. Using the InCise MLC system, the CK is able to deliver rectangular fields, allowing a more direct measurement of %dd(10cm) using conventional reference conditions. We define the PDD correction factor (CPDD) as the ratio of %dd(10 cm) measured using CK reference conditions to that measured using standard TG-51 reference conditions. Using four ionization chambers (A1SL, CC08, CC13, and A19), %dd(10 cm) is measured using a 6 cm fixed cone at 80 cm SSD and at 100 cm SSD using an effective 10 cm × 10 cm MLC-collimated field. These values are used to calculate CPDD, while the latter is used to directly calculate a kQ value. This direct kQ value is then compared to values determined using the BJR-25 method. Using the MLC system, this study demonstrates conversion between the %dd(10 cm) measured using CyberKnife reference conditions and TG-51 reference conditions. These values provide the means for derivation of a kQ curve as a function of direct measurements of %dd(10cm) using a 6 cm fixed-aperture collimating cone at 80 cm SSD.

Author supplied keywords

References Powered by Scopus

AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

1362Citations
N/AReaders
Get full text

A new formalism for reference dosimetry of small and nonstandard fields

479Citations
N/AReaders
Get full text

Addendum to the AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon beams

259Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Accurate machine-specific reference and small-field dosimetry for a self-shielded neuro-radiosurgical system

2Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Gersh, J. A., & Willett, B. (2015). Determination of kQ using MLC-collimated rectangular fields for absolute dosimetry of the CyberKnife. Journal of Applied Clinical Medical Physics, 16(6), 273–280. https://doi.org/10.1120/jacmp.v16i6.5720

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 2

100%

Readers' Discipline

Tooltip

Medicine and Dentistry 3

50%

Physics and Astronomy 3

50%

Save time finding and organizing research with Mendeley

Sign up for free