Real-time feedback improves chest compression quality in out-of-hospital cardiac arrest: A prospective cohort study

36Citations
Citations of this article
65Readers
Mendeley users who have this article in their library.

Abstract

Background: Current guidelines underline the importance of high-quality chest compression during cardiopulmonary resuscitation (CPR), to improve outcomes. Contrary to this many studies show that chest compression is often carried out poorly in clinical practice, and long interruptions in compression are observed. This prospective cohort study aimed to analyse whether chest compression quality changes when a real-time feedback system is used to provide simultaneous audiovisual feedback on chest compression quality. For this purpose, pauses in compression, compression frequency and compression depth were compared. Methods: The study included 292 out-of-hospital cardiac arrests in three consecutive study groups: first group, conventional resuscitation (no-sensor CPR); second group, using a feedback sensor to collect compression depth data without real-time feedback (sensor-only CPR); and third group, with real-time feedback on compression quality (sensor-feedback CPR). Pauses and frequency were analysed using compression artefacts on electrocardiography, and compression depth was measured using the feedback sensor. With this data, various parameters were determined in order to be able to compare the chest compression quality between the three consecutive groups. Results: The compression fraction increased with sensor-only CPR (group 2) in comparison with nosensor CPR (group 1) (80.1% vs. 87.49%; P < 0.001), but there were no further differences belonging compression fraction after activation of sensor-feedback CPR (group 3) (P = 1.00). Compression frequency declined over the three study groups, reaching the guideline recommendations (127.81 comp/min vs. 122.96 comp/min, P = 0.02 vs. 119.15 comp/min, P = 0.008) after activation of sensor-feedback CPR (group 3). Mean compression depth only changed minimally with sensor-feedback (52.49 mm vs. 54.66 mm; P = 0.16), but the fraction of compressions with sufficient depth (at least 5 cm) and compressions within the recommended 5-6 cm increased significantly with sensor-feedback CPR (56.90% vs. 71.03%; P = 0.003 and 28.74% vs. 43.97%; P < 0.001). Conclusions: The real-time feedback system improved chest compression quality regarding pauses in compression and compression frequency and facilitated compliance with the guideline recommendations. Compression depth did not change significantly after activation of the realtime feedback. Even the sole use of a CPR-feedback-sensor ("sensor-only CPR") improved performance regarding pauses in compression and compression frequency, a phenomenon known as the 'Hawthorne effect'. Based on this data real-time feedback systems can be expected to raise the quality level in some parts of chest compression quality.

Cite

CITATION STYLE

APA

Lakomek, F., Lukas, R. P., Brinkrolf, P., Mennewisch, A., Steinsiek, N., Gutendorf, P., … Bohn, A. (2020, February 1). Real-time feedback improves chest compression quality in out-of-hospital cardiac arrest: A prospective cohort study. PLoS ONE. Public Library of Science. https://doi.org/10.1371/journal.pone.0229431

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free