An assay for hepatitis B surface antigen (HBsAg) should reliably detect 0.2 μg/L, the lowest reported concentration in an asymptomatic blood donor. The difference between this concentration and the assay cutoff defines the analytical quality requirement in a total error format. The design of a statistical QC procedure is critically dependent on the precision of the assay. The precision of a developmental ELISA of HBsAg under study ranged from 17.5% to 9.6% for controls containing 0.07 to 1.50 μg/L, respectively. Use of one positive control with the 1(3s) QC rule provided an 85% chance of detecting a critical loss of assay sensitivity; use of two positive controls increased the chance of detecting critical loss of assay sensitivity to nearly 100%. These rules are based on the precision of this developmental assay, and must be developed individually for other assays. The development of the proposed QC procedures illustrates how quantitative QC can be provided for qualitative assays.
CITATION STYLE
Green IV, G. A., Neill Carey, R., Westgard, J. O., Carten, T., Shablesky, L., Achord, D., … Van Anh, L. E. (1997). Quality control for qualitative assays: Quantitative QC procedure designed to assure analytical quality required for an ELISA of hepatitis B surface antigen. Clinical Chemistry, 43(9), 1618–1621. https://doi.org/10.1093/clinchem/43.9.1618