Verification of cascade optical coherence tomography for freeform optics form metrology

  • Xu D
  • Wen Z
  • Coleto A
  • et al.
3Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Freeform optical components enable dramatic advances for optical systems in both performance and packaging. Surface form metrology of manufactured freeform optics remains a challenge and an active area of research. Towards addressing this challenge, we previously reported on a novel architecture, cascade optical coherence tomography (C-OCT), which was validated for its ability of high-precision sag measurement at a given point. Here, we demonstrate freeform surface measurements, enabled by the development of a custom optical-relay-based scanning mechanism and a unique high-speed rotation mechanism. Experimental results on a flat mirror demonstrate an RMS flatness of 14 nm (∼λ/44 at the He-Ne wavelength). Measurement on a freeform mirror is achieved with an RMS residual of 69 nm (∼λ/9). The system-level investigations and validation provide the groundwork for advancing C-OCT as a viable freeform metrology technique.

Cite

CITATION STYLE

APA

Xu, D., Wen, Z., Coleto, A. G., Pomerantz, M., Lambropoulos, J. C., & Rolland, J. P. (2021). Verification of cascade optical coherence tomography for freeform optics form metrology. Optics Express, 29(6), 8542. https://doi.org/10.1364/oe.413844