Development and validation of a RP-HPLC method for determination of nimodipine in sustained release tablets

2Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A rapid, sensitive, and reproducible reverse phase high performance liquid chromatographic (RP-HPLC) method with UV detector for the determination of nimodipine in sustained release tablets was developed. The method involved using a SinoChoom ODS-BP C18 reversed phase column (5 m, 4.6 mm × 200 mm) and mobile phase consisting of methanol-acetonitrile-water (35: 38: 27, v/v). The flow rate is 1.0 mL/min, the UV detector was operated at 237 nm, and the column was maintained at 25°C. The method was validated according to official compendia guidelines. The calibration curve of nimodipine for RP-HPLC method was linear over the range of 10-100 g/mL. The retention time was found at 7.50 min for nimodipine. The variation for interday and intraday assay was found to be less than 0.72%. The proposed RP-HPLC was proved to be suitable for the determination of nimodipine in sustained release tablets. © 2013 Xiaojun Shang et al.

Cite

CITATION STYLE

APA

Shang, X., Ma, S., & Li, Z. (2013). Development and validation of a RP-HPLC method for determination of nimodipine in sustained release tablets. Journal of Chemistry. https://doi.org/10.1155/2013/612082