Using a homogeneity test as weekly quality control on digital mammography units

5Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.
Get full text

Abstract

In the Netherlands a number of (screening) trials with digital mammography equipment have been started since 1999. In this paper results from the weekly QC procedure are given. It seems that the homogeneity test as described in the addendum to the European protocol is able to detect detector problems and flat field calibration problems. However, visual inspection remains necessary. For the CR system in the trials the homogeneity test did not find many problems. Either the homogeneity test is not effective and other tests might be more appropriate or this CR system does not have relevant image quality variations and therefore might not require weekly quality control. © Springer-Verlag Berlin Heidelberg 2006.

Cite

CITATION STYLE

APA

Van Engen, R. E., Swinkels, M. M. J., Oostveen, L. J., Geertse, T. D., & Visser, R. (2006). Using a homogeneity test as weekly quality control on digital mammography units. In Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics) (Vol. 4046 LNCS, pp. 259–265). Springer Verlag. https://doi.org/10.1007/11783237_36