We describe a multianalyte assay system for patient-side use comprising single-use plastic cartridges and a small monitor. Hemoglobin, glucose, and cholesterol can be simultaneously measured in 3 min in an unmeasured volume of blood. The sample is drawn by capillary action into four channels for delivery to assay-specific stacks containing a set of closely apposed layers. The distal layer is a membrane that acts as the optical surface for reflectance optics. For glucose and cholesterol assays, erythrocytes are removed by a fibrous filter layer and oxidase-peroxidase chemical reactions contained in the optical membrane generate a colored product. For hemoglobin measurement, blood is lysed by detergent contained in a porous disk. The amount of color reaching the optical membrane is measured by fiber optics. To ensure fail-safe operation, sensors verify sample sufficiency and degree of hemolysis. The assays perform comparably with laboratory methods.
CITATION STYLE
Sterling, B., Kiang, T., Subramanian, K., Saltman, M., Smart, W., Tsay, M., … Voss, F. (1992). Simultaneous patient-side measurement of hemoglobin, glucose, and cholesterol in finger-stick blood. Clinical Chemistry, 38(9), 1659–1664. https://doi.org/10.1093/clinchem/38.9.1658