Incidence and Factors Predicting Skin Burns at the Site of Indifferent Electrode during Radiofrequency Catheter Ablation of Cardiac Arrhythmias

7Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Radiofrequency catheter ablation (RFA) has become a mainstay for treatment of cardiac arrhythmias. Skin burns at the site of an indifferent electrode patch have been a rare, serious, and likely an underreported complication of RFA. The purpose of this study was to determine the incidence of skin burns in cardiac RFA procedures performed at one institution. Also, we wanted to determine the factors predicting skin burns after cardiac RFA procedures at the indifferent electrode skin pad site. Methods. A retrospective case control study was performed to compare the characteristics in patients who developed skin burns in a 2-year period. Results. Incidence of significant skin burns after RFA was 0.28% (6/2167). Four of the six patients were female and all were Caucasians. Four controls for every case were age and sex matched. Burn patients had significantly higher BMI, procedure time, and postprocedure pain, relative to control subjects (p < 0.05, one-tailed testing). No one in either group had evidence of dispersive pad malattachment. Conclusions. Our results indicate that burn patients had higher BMI and longer procedure times compared to control subjects. These findings warrant further larger studies on this topic.

Cite

CITATION STYLE

APA

Ibrahim, H., Finta, B., & Rind, J. (2016). Incidence and Factors Predicting Skin Burns at the Site of Indifferent Electrode during Radiofrequency Catheter Ablation of Cardiac Arrhythmias. Cardiology Research and Practice, 2016. https://doi.org/10.1155/2016/5265682

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free