Performance of epimyocardial leads in patients with a single ventricle circulation

1Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Cardiac pacing can be challenging after a Fontan operation, and limited data exist regarding strategies to plan these epimyocardial systems while minimizing the number of surgical procedures. Methods: A retrospective review of all our 47 patients (mean age 18 years, standard deviation 9 years) with a Fontan palliation who received an epimyocardial cardiac implantable electronic device (CIED) between 2002 and 2020 with regard to the stability of the epimyocardial lead parameters and the incidence of system revisions. Results: Over the last 18 years, 84 implantations or revisions of the epimyocardial CIED in 47 Fontan patients were performed. Mean age at time of the first implantation was 9.4 (range 0.28–29.3) years. Follow-up period ranges from 0.11 to 18.2 (mean 7.7, standard deviation 4.2) years. A total of 123 pacing leads were implanted of which 99 are still active. From 2010 triple lead cardiac resynchronization devices were used in 17 patients to better cope with lead problems. The initial pacing threshold of the leads inactivated during this study period proved significantly higher (mean 1.66 V) than in the "all leads" group (mean 1.27 V, p =.0005) or the group of the still active leads (mean 1.17 V, p =.00004). Conclusions: When implanted with a low pacing threshold, the bipolar epimyocardial electrodes show stable and good long-term results in young patients with a Fontan circulation. Resynchronization pacing systems and the prospective implantation of reserve leads may help to reduce the rate of resternotomies and provide a flexible concept to deal with lead failure.

Cite

CITATION STYLE

APA

Zartner, P. A., Mini, N., Vergnat, M., Momcilovic, D., Schneider, M. B., & Dittrich, S. (2021). Performance of epimyocardial leads in patients with a single ventricle circulation. PACE - Pacing and Clinical Electrophysiology, 44(5), 903–910. https://doi.org/10.1111/pace.14213

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free