Precision of 5 different keratometry devices

15Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.
Get full text

Abstract

To compare the precision among currently available keratometry devices. The corneal power was measured on two separate visits with the Nidek TonoRef II Autorefractor/Keratometer, the Zeiss IOLMaster 500, the Haag-Streit Lenstar LS 900, the Oculus Pentacam, and the Oculus Keratograph 4M. The precision was evaluated as the mean absolute intersession difference (MAD) between the corneal power measurements for each patient. Only the non-operated eye was included in the study. The Keratograph was found to have the highest MAD (0.215 D), which was significantly different from the other devices except for the IOLMaster. Nidek ARK had the lowest MAD (0.097 D), but this was not significant compared to Pentacam (0.124 D), Lenstar (0.132 D), or IOLMaster (0.140 D). Only one out of 29 patients had a precision difference exceeding 0.25 D with the Nidek ARK. Among the devices studied, the Nidek ARK was found to have the highest and the Keratograph was found to have to the lowest precision for the measurement of corneal power.

Cite

CITATION STYLE

APA

Laursen, J. V. N., Jeppesen, P., & Olsen, T. (2016). Precision of 5 different keratometry devices. International Ophthalmology, 36(1), 17–20. https://doi.org/10.1007/s10792-015-0069-3