The design of a pressure cell that compresses a cartilage specimen in one dimension within an imaging magnet is presented. One-dimensional projection images in a direction perpendicular to the articular surface of the cartilage specimen were used to generate a uniaxial confined deformation creep curve for normal and trypsin-degraded cartilage specimens during a continuous 0.690 MPa (100 psi) pressure application. The resulting curves are shown to fit a two time constant viscoelastic model well and also indicate that the elastic modulus of cartilage decreases and the deformation rate increases upon trypsin proteolysis. Furthermore, cartilage permeability is shown as a function of cartilage strain for both the normal and trypsin- degraded case. Several two-dimensional slice-selective images were collected both before and after 80 minutes of continuous compression. These images were used to evaluate the relative changes in the spin-lattice, T1, and spin- spin, T2, relaxation time constant maps for both normal and degraded cartilage specimens in response to compression. The results of this study demonstrate the utility of a novel, non-magnetic, cartilage compression device and also support the validity of a simple two-component rheological model of articular cartilage.
CITATION STYLE
Kaufman, J. H., Regatte, R. R., Bolinger, L., Kneeland, J. B., Reddy, R., & Leigh, J. S. (1999). A novel approach to observing articular cartilage deformation in vitro via magnetic resonance imaging. Journal of Magnetic Resonance Imaging, 9(5), 653–662. https://doi.org/10.1002/(SICI)1522-2586(199905)9:5<653::AID-JMRI6>3.0.CO;2-4