Measurement of oxygen consumption in children undergoing cardiac catheterization: Comparison between mass spectrometry and the breath-by-breath method

14Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Accurate measurement of oxygen consumption (VO2) is important to precise calculation of blood flow using the Fick equation. This study aimed to validate the breath-by-breath method (BBBM) of measuring oxygen consumption VO2 compared with respiratory mass spectroscopy (MS) for intubated children during cardiac catheterization. The study used MS and BBBM to measure VO2 continuously and simultaneously for 10 min in consecutive anesthetized children undergoing cardiac catheterization who were intubated with a cuffed endotracheal tube, ventilated mechanically, and hemodynamically stable, with normal body temperature. From 26 patients, 520 data points were obtained. The mean VO2 was 94.5 ml/min (95 % confidence interval [CI] 65.7-123.3 ml/min) as measured by MS and 91.4 ml/min (95 % CI 64.9-117.9 ml/min) as measured by BBBM. The mean difference in VO2 measurements between MS and BBBM (3.1 ml/min; 95 % CI -1.7 to +7.9 ml/min) was not significant (p = 0.19). The MS and BBBM VO2 measurements were highly correlated (R 2 = 0.98; P < 0.0001). Bland-Altman analysis showed good correspondence between MS and BBBM, with a mean difference of -3.01 and 95 % limits of agreement ranging from -26.2 to +20.0. The mean VO2 indexed to body surface area did not differ significantly between MS and BBBM (3.4 ml/min m2; 95 % CI -1.4 to 8.2; p = 0.162). The mean difference and limits of agreement were -3.8 ml/min m2 (range, -19.9 to 26.7). Both MS and BBBM may be used to measure VO2 in anesthetized intubated children undergoing cardiac catheterization. The two methods demonstrated excellent agreement. However, BBBM may be more suited to clinical use with children. © 2013 Springer Science+Business Media.

References Powered by Scopus

STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT

42119Citations
N/AReaders
Get full text

Errors in the measurement of cardiac output by thermodilution

259Citations
N/AReaders
Get full text

Determination of cardiac output by the Fick method, thermodilution, and acetylene rebreathing in pulmonary hypertension

215Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Hemodynamic assessment and acute pulmonary vasoreactivity testing in the evaluation of children with pulmonary vascular disease. Expert consensus statement on the diagnosis and treatment of paediatric pulmonary hypertension. The European Paediatric Pulmonary Vascular Disease Network, endorsed by ISHLT and DGPK

86Citations
N/AReaders
Get full text

Cardiac catheterization in children with pulmonary hypertensive vascular disease: Consensus statement from the Pulmonary Vascular Research Institute, Pediatric and Congenital Heart Disease Task Forces

51Citations
N/AReaders
Get full text

Cardiac Catheterization in Children with Pulmonary Hypertensive Vascular Disease

40Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Guo, L., Cui, Y., Pharis, S., Walsh, M., Atallah, J., Tan, M. W., … Adatia, I. (2014). Measurement of oxygen consumption in children undergoing cardiac catheterization: Comparison between mass spectrometry and the breath-by-breath method. Pediatric Cardiology, 35(5), 798–802. https://doi.org/10.1007/s00246-013-0855-5

Readers over time

‘17‘18‘20‘21‘22‘2400.511.52

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 2

50%

Professor / Associate Prof. 1

25%

Researcher 1

25%

Readers' Discipline

Tooltip

Medicine and Dentistry 3

60%

Biochemistry, Genetics and Molecular Bi... 1

20%

Sports and Recreations 1

20%

Save time finding and organizing research with Mendeley

Sign up for free
0