Beam hardening correction with an iterative scheme using an exact backward projector and a polychromatic forward projector

3Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.
Get full text

Abstract

In computed tomography (CT), reconstructions from cone-beam (CB) data acquired with a polychromatic X-ray device show so called beam hardening artifacts. Beam hardening artifacts are highly undesirable for a medical diagnosis, because details in the reconstructed image are severely disturbed or completely lost. In this work, we demonstrate the significant reduction of beam hardening artifacts by using an iterative reconstruction scheme which consists of a backward and a forward projector. For the backward projector, an exact reconstruction approach was used. The forward projector was extended by a polychromatic model to mimic a realistic X-ray device. The presented experiments use simulated CB data to restrict the evaluation to beam hardening artifacts. The discussion is focused on CB data acquired along a helical trajectory.

Cite

CITATION STYLE

APA

Bock, R., Hoppe, S., Scherl, H., & Hornegger, J. (2007). Beam hardening correction with an iterative scheme using an exact backward projector and a polychromatic forward projector. In Informatik aktuell (pp. 46–50). Kluwer Academic Publishers. https://doi.org/10.1007/978-3-540-71091-2_10

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free