Abstract

Purpose To demonstrate the effect of gradient nonlinearity and develop a method for correction of gradient nonlinearity artifacts in prospective motion correction (Mo-Co). Methods Nonlinear gradients can induce geometric distortions in magnetic resonance imaging, leading to pixel shifts with errors of up to several millimeters, thereby interfering with precise localization of anatomical structures. Prospective Mo-Co has been extended by conventional gradient warp correction applied to individual phase encoding steps/groups during the reconstruction. The gradient-related displacements are approximated using spherical harmonic functions. In addition, the combination of this method with a retrospective correction of the changes in the coil sensitivity profiles relative to the object (augmented sensitivity encoding (SENSE) reconstruction) was evaluated in simulation and experimental data. Results Prospective Mo-Co under gradient fields and coils sensitivity inconsistencies results in residual blurring, spatial distortion, and coil sensitivity mismatch artifacts. These errors can be considerably mitigated by the proposed method. High image quality with very little remaining artifacts was achieved after a few iterations. The relative image errors decreased from 25.7% to below 17.3% after 10 iterations. Conclusion The combined correction of gradient nonlinearity and sensitivity map variation leads to a pronounced reduction of residual motion artifacts in prospectively motion-corrected data.

References Powered by Scopus

Cubic Convolution Interpolation for Digital Image Processing

3220Citations
N/AReaders
Get full text

Advances in sensitivity encoding with arbitrary k-space trajectories

1015Citations
N/AReaders
Get full text

Motion correction with PROPELLER MRI: Application to head motion and free-breathing cardiac imaging

789Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Pros and cons of ultra-high-field MRI/MRS for human application

378Citations
N/AReaders
Get full text

Motion correction in MRI of the brain

168Citations
N/AReaders
Get full text

Prospective motion correction in functional MRI

90Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Yarach, U., Luengviriya, C., Danishad, A., Stucht, D., Godenschweger, F., Schulze, P., & Speck, O. (2015). Correction of gradient nonlinearity artifacts in prospective motion correction for 7T MRI. Magnetic Resonance in Medicine, 73(4), 1562–1569. https://doi.org/10.1002/mrm.25283

Readers over time

‘14‘15‘16‘17‘18‘19‘20‘21‘22‘23‘24036912

Readers' Seniority

Tooltip

Researcher 18

56%

PhD / Post grad / Masters / Doc 11

34%

Professor / Associate Prof. 2

6%

Lecturer / Post doc 1

3%

Readers' Discipline

Tooltip

Physics and Astronomy 9

33%

Engineering 9

33%

Neuroscience 5

19%

Medicine and Dentistry 4

15%

Article Metrics

Tooltip
Social Media
Shares, Likes & Comments: 37

Save time finding and organizing research with Mendeley

Sign up for free
0