This article evaluates the potential of capacitive measurements using flexible electrodes to access various physical quantities. These electrodes are made of a thin metallic film, typical thickness 0.2 μm, evaporated on a plastic substrate. Their large flexibility enables them to be mounted in complex geometries such as curved surfaces. In the configuration of planar condensers, using a very sensitive commercial capacitive bridge and a three-terminal measurement method, several measurements are presented. A relative resolution of 10-8 for the thermal expansion of samples is obtained at low temperature in a differential configuration. The same technique adopted for pressure gauge measurements at low temperature led to a typical 0.1 Pa resolution over a dynamic range of 104 Pa. In the configuration of interleaved electrodes, condensers have been used to measure wetting by either bulk liquid helium or by thin continuous helium films in a cylindrical pipe. Both experimental and numerical evidence is provided, showing that the close proximity of a reference ground potential significantly increases the relative sensitivity to fluid wetting. Further, interleaved electrodes can be used to access both the area that is covered by a liquid film but also to determine the thickness of this film, provided it is comparable to the periodicity of the electrode pattern. © 2007 American Institute of Physics.
CITATION STYLE
Thibault, P., Diribarne, P., Fournier, T., Perraud, S., Puech, L., Wolf, P. E., … Vallcorba, R. (2007). On the design of capacitive sensors using flexible electrodes for multipurpose measurements. Review of Scientific Instruments, 78(4). https://doi.org/10.1063/1.2721406