A high-throughput, fully automated antigen detection test for SARS-CoV-2 is a viable alternative to reverse-transcription polymerase chain reaction (RT-qPCR) for mass screening during outbreaks. In this study, we compared RT-qPCR for viral load and the VITROS® SARS-CoV-2 Antigen Test with reference to the results of the LUMIPULSE® SARS-CoV-2 Ag Test. Of 128 nasopharyngeal swab specimens taken from patients suspected of being infected with SARS-CoV-2, 49 were positive and 79 were negative according to RT-qPCR. Consistent dose-dependent detection with VITROS® assay was successfully achieved when using nasopharyngeal swab specimens with Ct values of 32.0 or lesser, whereas the CLEIA-based LUMIPULSE® assay was able to detect lower viral loads compared with the VITROS® assay. Our results show that the performance of the VITROS® assay was satisfactory for the diagnosis of contagious COVID-19 patients in the clinical setting. Highlights The performance of the VITROS® SARS-CoV-2 Antigen Test was sufficient for the diagnosis of contagious COVID-19. This test showed high sensitivity and specificity in the detection of SARS-CoV-2 in samples with a Ct value of 32 or less.
CITATION STYLE
Matsuzaki, N., Orihara, Y., Kodana, M., Kitagawa, Y., Matsuoka, M., Kawamura, R., … Maeda, T. (2021). Evaluation of a chemiluminescent enzyme immunoassay-based high-throughput SARS-CoV-2 antigen assay for the diagnosis of COVID-19: The VITROS® SARS-CoV-2 Antigen Test. Journal of Medical Virology, 93(12), 6778–6781. https://doi.org/10.1002/jmv.27153