Evaluation of the new digital goldmann applanation tonometer for measuring intraocular pressure

3Citations
Citations of this article
15Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose. To compare a new digital Goldmann applanation tonometer (dGAT) that measures intraocular pressure (IOP) in 0.1 mmHg increments to a standard Goldmann applanation tonometer (sGAT). Methods. This study included 116 eyes of 60 subjects. A single examiner first measured IOP in triplicate using either sGAT or dGAT, which was randomly chosen. After a 5-minute interval, the next set of three consecutive IOP was measured using the other GAT. Results. The mean IOP measured with sGAT was 16.27 ± 6.68 mmHg and 16.35 ± 6.69 mmHg with dGAT. Pearson's correlation coefficient was 0.998 (P < 0.01). The subjects were divided into three groups based on the mean IOP: IOP < 14 mmHg, 14-20 mmHg, or >20 mmHg. The Pearson's correlation coefficient within each group was 0.935, 0.972, and 0.997 (P < 0.01), respectively. The difference within the three consecutive IOP measurements (maximum-minimum) for dGAT (0.72 ± 0.34 mmHg) was significantly smaller than those with sGAT (0.92 ± 0.42 mmHg, P < 0.01). Even in patients with equal IOP (zero left-right difference) with sGAT (n = 30), dGAT detected IOP differences between the left and right eyes (0.47 ± 0.31 mmHg). Conclusion. Compared to sGAT, dGAT measurements are highly reproducible and less variable. © 2014 Yuta Sakaue et al.

Cite

CITATION STYLE

APA

Sakaue, Y., Ueda, J., Seki, M., Tanaka, T., Togano, T., Yoshino, T., & Fukuchi, T. (2014). Evaluation of the new digital goldmann applanation tonometer for measuring intraocular pressure. Journal of Ophthalmology, 2014. https://doi.org/10.1155/2014/461681

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free