Comparison of in vivo and in vitro models to evaluate pulp temperature rise during exposure to a Polywave® LED light curing unit

15Citations
Citations of this article
41Readers
Mendeley users who have this article in their library.

Abstract

Objectives: To measure and compare in vivo and in vitro pulp temperature (PT) increase (ΔTEMP) over baseline, physiologic temperature using the same intact upper premolars exposed to the same Polywave® LED curing light. Methodology: After local Ethics Committee approval (#255,945), local anesthesia, rubber dam isolation, small occlusal preparations/minute pulp exposure (n=15) were performed in teeth requiring extraction for orthodontic reasons. A sterile probe of a temperature measurement system (Temperature Data Acquisition, Physitemp) was placed within the pulp chamber and the buccal surface was sequentially exposed to a LED LCU (Bluephase 20i, Ivoclar Vivadent) using the following exposure modes: 10-s low or high, 5-s Turbo, and 60-s high. Afterwards, the teeth were extracted and K-type thermocouples were placed within the pulp chamber through the original access. The teeth were attached to an assembly simulating the in vivo environment, being similarly exposed while real-time temperature (°C) was recorded. ΔTEMP values and time for temperature to reach maximum (ΔTIME) were subjected to two-way ANOVA and Bonferroni's post-hoc tests (pre-set alpha 0.05). Results: Higher ΔTEMP was observed in vitro than in vivo. No significant difference in ΔTIME was observed between test conditions. A significant, positive relationship was observed between radiant exposure and ΔTEMP for both conditions (in vivo: r2=0.917; p<0.001; in vitro: r2=0.919; p<0.001). Conclusion: Although the in vitro model overestimated in vivo PT increase, in vitro PT rise was close to in vivo values for clinically relevant exposure modes.

Cite

CITATION STYLE

APA

Runnacles, P., Arrais, C. A. G., Maucoski, C., Coelho, U., DE GOES, M. F., & Rueggeberg, F. A. (2019). Comparison of in vivo and in vitro models to evaluate pulp temperature rise during exposure to a Polywave® LED light curing unit. Journal of Applied Oral Science, 27. https://doi.org/10.1590/1678-7757-2018-0480

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free