Monte Carlo modeling of the Elekta Versa HD and patient dose calculation with EGSnrc/BEAMnrc

2Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

Abstract

Introduction: Numerous studies have proven the Monte Carlo method to be an accurate means of dose calculation. Although there are several commercial Monte Carlo treatment planning systems (TPSs), some clinics may not have access to these resources. We present a method for routine, independent patient dose calculations from treatment plans generated in a commercial TPS with our own Monte Carlo model using free, open-source software. Materials and methods: A model of the Elekta Versa HD linear accelerator was developed using the EGSnrc codes. A MATLAB script was created to take clinical patient plans and convert the DICOM RTP files into a format usable by EGSnrc. Ten patients’ treatment plans were exported from the Monaco TPS to be recalculated using EGSnrc. Treatment simulations were done in BEAMnrc, and doses were calculated using Source 21 in DOSXYZnrc. Results were compared to patient plans calculated in the Monaco TPS and evaluated in Verisoft with a gamma criterion of 3%/2 mm. Results: Our Monte Carlo model was validated within 1%/1-mm accuracy of measured percent depth doses and profiles. Gamma passing rates ranged from 82.1% to 99.8%, with 7 out of 10 plans having a gamma pass rate over 95%. Lung and prostate patients showed the best agreement with doses calculated in Monaco. All statistical uncertainties in DOSXYZnrc were less than 3.0%. Conclusion: A Monte Carlo model for routine patient dose calculation was successfully developed and tested. This model allows users to directly recalculate DICOM RP files containing patients’ plans that have been exported from a commercial TPS.

References Powered by Scopus

A technique for the quantitative evaluation of dose distributions

2397Citations
N/AReaders
Get full text

Dosimetric considerations for patients with HIP prostheses undergoing pelvic irradiation. Report of the AAPM Radiation Therapy Committee Task Group 63

227Citations
N/AReaders
Get full text

Large efficiency improvements in BEAMnrc using directional bremsstrahlung splitting

188Citations
N/AReaders
Get full text

Cited by Powered by Scopus

A Monte Carlo model of an agility head for a 6 MV Elekta photon beam

0Citations
N/AReaders
Get full text

Monte Carlo simulation and comparison for the penumbra characteristics of flattening filter and flattening filter free photon beams from medical electron linear accelerator

0Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Paschal, H. M., Kabat, C. N., Papaconstadopoulos, P., Kirby, N. A., Myers, P. A., Wagner, T. D., & Stathakis, S. (2022). Monte Carlo modeling of the Elekta Versa HD and patient dose calculation with EGSnrc/BEAMnrc. Journal of Applied Clinical Medical Physics, 23(9). https://doi.org/10.1002/acm2.13715

Readers over time

‘22‘23‘24‘250481216

Readers' Seniority

Tooltip

Researcher 3

60%

Lecturer / Post doc 1

20%

PhD / Post grad / Masters / Doc 1

20%

Readers' Discipline

Tooltip

Medicine and Dentistry 2

40%

Agricultural and Biological Sciences 1

20%

Engineering 1

20%

Computer Science 1

20%

Save time finding and organizing research with Mendeley

Sign up for free
0